Die u:cris Detailansicht:
Magnetoresistance of the new ceramic "Cernox" thermometer from 4.2 K to 300 K in magnetic fields up to 13 T
- Autor(en)
- G. Heine, W. Lang
- Abstrakt
Measurements of the resistance and the magnetoresistance of a new ceramic thin-film temperature sensor (Cernox-1080, Lake Shore Cryotronics Inc.) are reported for magnetic fields up to 13 T in the temperature range 4.2 to 300 K. The zero-field resistance changes monotonically over more than three orders of magnitude with a negative temperature coefficient. The magnetoresistance is small at temperatures above 10 K and exhibits partially compensating positive and negative components. The magnetic-field induced temperature reading error at 13 T is about 2% at 4.2 K and less than 0.3% at 10 K and above. In the temperature range 80 to 300 K, the magnetoresistance can be approximated by a fourth-order polynomial, allowing for an estimation and correction of the apparent temperature reading error of the Cernox thermometer in high magnetic fields.
- Organisation(en)
- Elektronische Materialeigenschaften
- Journal
- Cryogenics
- Band
- 38
- Seiten
- 377-379
- Anzahl der Seiten
- 3
- ISSN
- 0011-2275
- DOI
- https://doi.org/10.1016/S0011-2275(97)00130-6
- Publikationsdatum
- 1998
- Peer-reviewed
- Ja
- ÖFOS 2012
- 103018 Materialphysik
- Schlagwörter
- Link zum Portal
- https://ucrisportal.univie.ac.at/de/publications/f8339a55-96bb-49cd-99d0-91d1dfea488d