A Medium Flow, High-Resolution Vienna DMA Running in Recirculating Mode
- Autor(en)
- Gerhard Steiner, Michel Attoui, Daniela Wimmer, Georg Reischl
- Abstrakt
In this work we present the latest design of the Vienna type Differential Mobility Analyzer (DMA). The so-called UDMA is specially designed for high-resolution mobility measurements for the particle size range between 1 and 5 nm and has been optimized to be combined with a mass spectrometer to provide a well-defined mobility fraction of molecular clusters. As the UDMA is designed for this narrow size range it features high-resolution mobility measurements already at flow rates of approximately 700 L/min, marked as medium flow conditions. This readily allows operating the UDMA in a recirculating sheath air flow system with gas cleaning filters, as the control over the purity of the sheath air is crucial for operating the UDMA inline with a mass spectrometer or other carrier gas sensible devices.